Customer reported receiving two erroneous aptt-ss results on two different patients on their acl top 300 cts instrument with hemosil synthasil aptt reagent.Per the customer, their morning qc passed, but the second shift failed.All patient samples tested between those two time periods were repeated after placing fresh reagent on board the instrument and obtaining control values in range at 15:30.The results obtained at that point correlated with the values obtained before the failed qc with the exception of the two aptt results referenced above.Both patients were treated with heparin based on the original results.The physician was alerted and further testing performed to ensure patients were back in therapeutic range.The customer confirmed no impact on the treated patients.
|
Recovery of qc in range at 15:30pm, after replacing new aptt reagents, suggest that this incident was associated with contaminated aptt reagents.Due to many unknown factors of reagent handling on site and this being the only incident reported of this occurrence, qa eval associates this as an isolated random error most likely due to reagent handling.Based on this analysis no remedial action is required.
|