Model Number D314TRG
Device Problems Compatibility Problem (2960); Output Problem (3005)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/20/2015
Event Type  malfunction  
Event Description
It was reported that short interval counts (sic) were observed with the right ventricular (rv) lead.In addition, the rv electrogram exhibited missing data.It was further reported that the implanting physician intentionally placed the left ventricular (lv) lead into the device's rv port.The rv lead and device remain in use.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.Concomitant medical products: 5076-45 lead, implanted: (b)(6) 2008; 1156t st jude competitor lead, implanted: (b)(6) 2008.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTECTA XT CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
ceiba norte industrial park, r
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key4830407
MDR Text Key5842464
Report Number3004209178-2015-10931
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 04/20/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/10/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/14/2013
Device Model NumberD314TRG
Device Catalogue NumberD314TRG
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received04/20/2015
Date Device Manufactured01/25/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
6949-65 LEAD
Patient Age00073 YR
-
-