• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION OCTRODE LEAD KIT, 60CM LENGTH; SCS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION OCTRODE LEAD KIT, 60CM LENGTH; SCS LEAD Back to Search Results
Model Number 3186
Device Problems Material Fragmentation (1261); Kinked (1339); Material Twisted/Bent (2981)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/20/2015
Event Type  Injury  
Event Description
It was reported, the patient underwent a permanent implant procedure on (b)(6) 2015.During the procedure, the hub of the stylet broke off when the doctor was trying to remove the stylet from the lead.X-rays were taken and showed the lead was kinked and twisted.As a result, the doctor removed and replaced the lead.The issue was resolved by the replacement lead.The procedure was extended by 1.5 hours.
 
Manufacturer Narrative
(b)(4).Sjm has limited information related to the patient¿s medical history and is unable to form an opinion as to the relevancy of the patient¿s history to the event reported.Sjm defers to the patient¿s physician regarding medical history.
 
Manufacturer Narrative
Sjm has limited information related to the patient¿s medical history and is unable to form an opinion as to the relevancy of the patient¿s history to the event reported.Sjm defers to the patient¿s physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCTRODE LEAD KIT, 60CM LENGTH
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key4837884
MDR Text Key17211263
Report Number1627487-2015-07293
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative,company representative
Reporter Occupation Other
Type of Report Initial
Report Date 07/20/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2017
Device Model Number3186
Device Lot Number4990837
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/01/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/20/2015
Initial Date FDA Received06/11/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/24/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age75 YR
-
-