• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. RUBY COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. RUBY COIL; HCG, KRD Back to Search Results
Catalog Number RBY4C0315
Device Problems Component Falling (1105); Contamination (1120)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/12/2015
Event Type  malfunction  
Manufacturer Narrative
The pet lock was intact.The pusher assembly was kinked approximately 8.5 cm from the proximal end.The complaint has been evaluated.The complaint indicated that during the procedure, the ruby coil was dropped on the floor and was not used further.Evaluation of the returned device revealed the pusher assembly was kinked.This type of damage typically occurs due to improper handling during use.If the device is dropped on the floor during use, damage such as this may occur.These devices are 100% visually inspected and functionally tested during process inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event Description
The patient was undergoing a coil embolization procedure using ruby coils.During the procedure, the ruby coil was dropped on the ground, rendering it unsterile.The procedure continued successfully using a new ruby coil.
 
Manufacturer Narrative
Conclusion: the device has been returned and the investigation results are pending.A follow up mdr will be submitted upon completion of the device investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL
Type of Device
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer (Section G)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
kathleen kidd
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key4842314
MDR Text Key13491704
Report Number3005168196-2015-00587
Device Sequence Number1
Product Code HCG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K120330
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/13/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue NumberRBY4C0315
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/27/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/10/2015
Initial Date FDA Received06/12/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/15/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-