Catalog Number 8065990836 |
Device Problem
Unintended System Motion (1430)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/24/2015 |
Event Type
malfunction
|
Event Description
|
A surgeon reported to company representative that during a lasik procedure the patient bed was moving downward.Reporter indicated the surgeon had to stop and refocus until the uncontrollable bed movement stopped.Reporter relayed the procedure was completed and there was no patient harm; however the procedure time was delayed.
|
|
Manufacturer Narrative
|
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The device history records (dhr) for the device was reviewed.The associated device was released based on company acceptance criteria.The manufacturer internal reference number is: (b)(4).
|
|
Manufacturer Narrative
|
During the onsite visit the fse (field service engineer) replaced the z-motor of the patient bed and performed adjustment.The failure analysis shows the problem cannot be reproduced.The motor worked fine during complete testing and with the maximum of the specified load.The root cause could not be determined conclusively as the failure analysis shows no problem with the sample.Based on information received following submission of the initial report, this event does not meet criteria for reporting as a device malfunction.(b)(4).
|
|
Event Description
|
Upon additional follow up, reporter indicated the reported issue occured prior to the laser firing only, and not during the laser treatment.
|
|
Search Alerts/Recalls
|