ST. JUDE MEDICAL PUERTO RICO, INC. (CS) SJM MASTERS SERIES VALVE EXPANDED CUFF; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 23MECJ-502
Device Problem Perivalvular Leak (1457)
Patient Problem Calcium Deposits/Calcification (1758)
Event Date 06/08/2015
Event Type  Injury  
Event Description
On (b)(6) 2015, a 19mm regent mechanical heart valve was implanted in the aortic position and a 23mm masters series mechanical heart valve was implanted in the mitral position.On june 8, re-do surgery was performed on the mitral valve due to paravalvular leakage (pvl) secondary to a suspected calcified and torn native annulus.Two additional sutures with felt pledgets were required for the mechanical mitral valve which resolved the pvl.The mitral valve remains implanted.As a result of the pledgets, the aortic valve area became narrower and it was necessary to explant the 29mm regent valve and implant a smaller 17mm regent valve.
 
Manufacturer Narrative
(b)(4).The results of the investigation are inconclusive since the device was not returned for analysis.Our investigation was limited to the review of the device history record, which showed that each manufacturing and inspection operation was performed and indicated complete in accordance with sjm specifications and procedures.Based on the information received, the cause of the reported incident could not be conclusively determined.This report is related to mdr 2648612-2015-00012 for the 19mm regent valve that was explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM MASTERS SERIES VALVE EXPANDED CUFF
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, INC. (CS)
p.o. box 998
lot 20 b st.
caguas, puerto rico 0072 5
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC. (CS)
p.o. box 998
lot 20 b st.
caguas, puerto rico 0072 5
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key4853897
MDR Text Key19986956
Report Number2648612-2015-00013
Device Sequence Number1
Product Code LWQ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 06/11/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/18/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date02/11/2019
Device Model Number23MECJ-502
Device Catalogue Number23MECJ-502
Device Lot Number4466753
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/11/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/12/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-