Catalog Number 530.605
Device Problem Loss of Power (1475)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/18/2015
Event Type  malfunction  
Event Description
It was reported that during an unspecified surgical procedure, it was discovered that the battery reamer/drill device stopped during reaming.There were no delays to the planned surgical procedure as a spare device was available for use.The reporter indicated that the procedure was completed without further incident.There was patient involvement reported.There were no reports of injuries, medical intervention, or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the device passed all operational specifications.Therefore, the reported condition was not duplicated or confirmed.An assignable root cause was not determined.However, during evaluation, it was observed that there was liquid internally on electronic and mechanical components.It was determined that this was due to immersion during the cleaning process.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY REAMER/DRILL
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
SYNTHES OBERDORF
eimattstrasse 3
oberdorf CH44 36
SZ  CH4436
Manufacturer (Section G)
SYNTHES OBERDORF
eimattstrasse 3
oberdorf CH44 36
SZ   CH4436
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key4857280
MDR Text Key5811594
Report Number8030965-2015-11028
Device Sequence Number1
Product Code GEY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Nurse
Type of Report Initial
Report Date 05/20/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number530.605
Device Lot Number4963
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/20/2015
Initial Date FDA Received06/19/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/26/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age92 YR
-
-