It was reported to covidien on (b)(6) 2015 that a customer had an issue with a dialysis catheter.The customer reports the catheter was blocked after the first treatment and the patient was unable to do the next dialysis treatment.After taking out the catheter from the patient, the customer discovered there was a blood clot at the side slot, therefore causing the blockage of the catheter.The customer followed normal practice, infused 2ml heparin 5000iu in each lumen as locking solution to prevent catheter clot.
|
The device history record was reviewed and indicated that the product was released accomplishing all quality standards.There were no samples received for evaluation.Because a sample was not available for evaluation, a root cause analysis could not be conducted to determine the root cause of this reported issue.If samples are received at a later date, this complaint will be re-opened and the investigation continued.A corrective action is not applicable at this time.It must be noted that in-process controls (such as personnel training, incoming quality acceptance testing for raw material, in process visual inspection and visual acceptance sampling).Additionally, the devices are inspected per extension assembly procedure in order to discard occlusion.This complaint will be used for tracking and trending purposes.
|