ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) UNIFY ASSURA CRT-D RF HV; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3357-40Q
Device Problems Difficult to Interrogate (1331); Delayed Charge Time (2586)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/26/2015
Event Type  Injury  
Event Description
It was reported that the asymptomatic patient presented in clinic after receiving an alert for the capacitor charge time limit having been reached.During attempted capacitor maintenances in-clinic the device exhibited loss of telemetry and continued capacitor charge time outs.The device was explanted and replaced.
 
Manufacturer Narrative
(b)(4).
 
Manufacturer Narrative
The reported extended charge time was confirmed in the laboratory.The device was tested on the bench as well as using automated test equipment, and no anomaly was found.The capacitors were returned to the manufacturer for further analysis and an internal capacitor anomaly was found to be the cause of the reported extended charge time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY ASSURA CRT-D RF HV
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
jaime chavez
645 almanor avenue
sunnyvale, CA 94085
8184934022
MDR Report Key4863912
MDR Text Key5808938
Report Number2938836-2015-27284
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative,company representati
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/26/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2015
Device Model NumberCD3357-40Q
Device Lot Number4385209
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/29/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/26/2015
Initial Date FDA Received06/23/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received07/16/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/12/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age76 YR
-
-