U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION LAMITRODE TRIPOLE 16 LEAD, 60 CM; SCS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION LAMITRODE TRIPOLE 16 LEAD, 60 CM; SCS LEAD Back to Search Results
Model Number 3219
Device Problems Device Operates Differently Than Expected (2913); Improper Device Output (2953)
Patient Problems Therapeutic Effects, Unexpected (2099); Inadequate Pain Relief (2388)
Event Date 05/15/2015
Event Type  Injury  
Event Description
It was reported, the patient has been receiving ineffective stimulation.It was also reported, the patient has been receiving stimulation in an unintended area.In turn, the patient may undergo x-rays and surgical intervention.
 
Manufacturer Narrative
(b)(4).Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Manufacturer Narrative
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Event Description
Follow-up revealed x-rays were taken and no anomalies were found.Reprogramming to provide effective therapy was unsuccessful.Per the manufacturer's device record database, the patient's lead was explanted and replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAMITRODE TRIPOLE 16 LEAD, 60 CM
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key4872941
MDR Text Key17540127
Report Number1627487-2015-05304
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,other
Reporter Occupation Other
Type of Report Initial
Report Date 07/27/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/28/2013
Device Model Number3219
Device Lot Number3443850
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/04/2015
Initial Date FDA Received06/26/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/18/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age69 YR
-
-