Model Number PM2240
Device Problem Programming Issue (3014)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/07/2015
Event Type  malfunction  
Event Description
It was reported that during routine follow-up, it was observed that retrograde conduction had resulted in automated mode switch episodes.No patient symptoms were reported.Device reprogramming was performed and the retrograde conduction was suppressed.
 
Manufacturer Narrative
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSURITY RF DR
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
jaime chavez
15900 valley view court
sylmar, CA 91342
8184934022
MDR Report Key4873486
MDR Text Key22249131
Report Number2017865-2015-26403
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Nurse
Type of Report Initial
Report Date 04/07/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/27/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/29/2016
Device Model NumberPM2240
Device Lot Number4757598
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/07/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/22/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age78 YR
-
-