• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOLOGIC INC. THINPREP 5000 PROCESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOLOGIC INC. THINPREP 5000 PROCESSOR Back to Search Results
Device Problem No Display/Image (1183)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2015
Event Type  malfunction  
Event Description
A customer in (b)(6) initially reported their thinprep 5000 processor failed to read the barcode and requested service from hologic inc.Hologic dispatched a field service engineer (fse) to investigate the incident.The fse determined the instrument mismatched the slide and vial id.There was no delay in patient diagnosis, no misdiagnosis and the patient was not recalled as a result of this incident.The sample was reprocessed with the correct sample identifier etched on the slide.This is considered a reportable event since the thinprep 5000 processor did not perform as intended and this could potentially lead to a misdiagnosis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THINPREP 5000 PROCESSOR
Manufacturer (Section D)
HOLOGIC INC.
marlborough MA
Manufacturer Contact
eva ,axwe;;
250 campus drive
marlborough, MA 01752
MDR Report Key4873872
MDR Text Key6059632
Report Number1222780-2015-00116
Device Sequence Number1
Product Code MKQ
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Unknown
Reporter Occupation Other
Type of Report Initial
Report Date 06/25/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 05/29/2015
Initial Date FDA Received06/25/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-