• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH MCREYNOLDS IMPACTOR ADAPTER; INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH MCREYNOLDS IMPACTOR ADAPTER; INSTRUMENT Back to Search Results
Catalog Number 6260-4-090
Device Problems Break (1069); Material Integrity Problem (2978)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/03/2015
Event Type  malfunction  
Event Description
Surgeon attempted to separate the body from the stem using the 6278-9-070 body/stem separator, lot tace913.The tip of this broke off inside the cone body.Surgeon was able to remove broken pieces from inside the cone body using forceps.Surgeon then attempted to remove entire construct using 6260-4-090 mcreynolds distal stem adaptor lot tacl506, on the slap hammer.The thread of the mcreynolds distal stem adaptor snapped off inside the implant.
 
Manufacturer Narrative
A supplemental report will be submitted upon completion of the investigation.
 
Manufacturer Narrative
The event was confirmed by the return of the fractured device.Visual inspection: the device was returned with the threaded tip broken from the device at the first thread.The rest of the device showed normal signs of use after 5 years in the field.The root cause of the reported event was determined by a material analysis sme to be bending overload.No further investigation for this event is possible at this time as insufficient information was received by stryker orthopaedics.If additional information becomes available, this investigation will be reopened.
 
Event Description
Surgeon attempted to separate the body from the stem using the 6278-9-070 body/stem separator, lot tace913.The tip of this broke off inside the cone body.Surgeon was able to remove broken pieces from inside the cone body using forceps.Surgeon then attempted to remove entire construct using 6260-4-090 mcreynolds distal stem adaptor lot tacl506, on the slap hammer.The thread of the mcreynolds distal stem adaptor snapped off inside the implant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCREYNOLDS IMPACTOR ADAPTER
Type of Device
INSTRUMENT
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
william hanna
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key4874456
MDR Text Key21800788
Report Number0002249697-2015-02125
Device Sequence Number1
Product Code MEH
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K122853
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative,company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/03/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number6260-4-090
Device Lot NumberTACL506
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/03/2015
Initial Date FDA Received06/29/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received10/29/2015
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/24/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age62 YR
-
-