STRYKER MEDICAL LONDON DBA CHG HOSPITAL BEDS SPIRIT SELECT; BED, AC-POWERED ADJUSTABLE HOSPITAL Back to Search Results
Catalog Number 5700000000
Device Problems False Reading From Device Non-Compliance (1228); Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/13/2015
Event Type  malfunction  
Event Description
It was reported that the scale was inaccurate due to the patient left foot end load cell not functioning.No patient was affected and no adverse consequence or clinically relevant delay in treatment was reported.
 
Manufacturer Narrative
Chg hospital beds, inc.Was acquired by stryker on jan.2, 2015.This medwatch is being submitted late because it was identified during the integration and remediation activities initiated by stryker medical in conjunction with this acquisition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPIRIT SELECT
Type of Device
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
STRYKER MEDICAL LONDON DBA CHG HOSPITAL BEDS
1020 adelaide st. s.
N6E 1R6
CA  N6E 1R6
Manufacturer (Section G)
STRYKER MEDICAL LONDON DBA CHG HOSPITAL BEDS
1020 adelaide st. s.
N6E 1R6
CA   N6E 1R6
Manufacturer Contact
rita moffitt
3800 east centre avenue
portage, MI 49002
2693292100
MDR Report Key4878766
MDR Text Key6057246
Report Number3006433555-2015-00237
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/13/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number5700000000
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 03/13/2015
Initial Date FDA Received06/30/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-