Device Problems Leak/Splash (1354); Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/12/2015
Event Type  malfunction  
Event Description
It was reported that a system error 2240 alarm (air in set/line) occurred on the homechoice device during drain four of six in peritoneal dialysis.The patient was connected at the time of the alarm.The technical service representative assisted with proper disconnect procedures and advised the nurse to cycle the power on the device to clear the alarm.The nurse reported that the supply bag was leaking.The patient would perform a manual drain.There was no patient injury or medical intervention indicated at the time of the report.No additional information is available.
 
Manufacturer Narrative
(b)(4).An alarm indicative of a potential malfunction of the disposable cassette was reported.The device was not returned and the lot number is unknown; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key4881338
MDR Text Key6260311
Report Number1416980-2015-27301
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 06/12/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/01/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/12/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
HOMECHOICE
-
-