Device Problems Kinked (1339); Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/14/2015
Event Type  malfunction  
Event Description
It was reported that a homechoice device experienced a system error 2240 (air in line/set) alarm.The patient was connected at the time of the alarm.This occurred during initial drain of peritoneal dialysis therapy.During troubleshooting, it was discovered the patient line was not properly primed due to a kink in the tubing.The technical service representative reviewed proper procedures with the patient and advised them restart therapy using all new supplies.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
(b)(4).The device was not returned and the lot number was unknown; therefore, a device analysis could not be completed.However, the assignable cause was determined to be kinked tubing.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key4883715
MDR Text Key16561022
Report Number1416980-2015-27411
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 06/14/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/14/2015
Initial Date FDA Received07/01/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
HOMECHOICE
-
-