• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECKMAN COULTER COULTER LH 750 HEMATOLOGY ANALYZER; COUNTER, DIFFERENTIAL CELL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECKMAN COULTER COULTER LH 750 HEMATOLOGY ANALYZER; COUNTER, DIFFERENTIAL CELL Back to Search Results
Catalog Number 6605731
Device Problems Thermal Decomposition of Device (1071); Smoking (1585)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/15/2015
Event Type  malfunction  
Event Description
The customer reported that there was a burning smell coming from a coulter lh 750 hematology analyzer, and a service visit was requested.There was smoke reported by the customer.The fire department was not called and there was no death or injury as a result.There was no medical attention required by any operator.Erroneous patient results were not generated and there was no change or effect to patient treatment in connection to the event.There was no impact to patient results and controls.
 
Manufacturer Narrative
A field service engineer (fse) reported an electrical failure in the power supply that caused flame damage to some components in the power supply area.Images were provided that show the damage caused by the power supply.The damaged power supply was replaced.The repairs were verified per established service procedures.(b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COULTER LH 750 HEMATOLOGY ANALYZER
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s kraemer blvd
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th avenue
miami FL 33196
Manufacturer Contact
miranda holland
11800 sw 147th avenue, 32-s08
miami, FL 33196-2031
3053802031
MDR Report Key4888654
MDR Text Key6537711
Report Number1061932-2015-01077
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeCI
PMA/PMN Number
K061574
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/15/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/02/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number6605731
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/15/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-