DEPUY SYNTHES POWER TOOLS SMALL BATTERY DRIVE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 532.010 |
Device Problem
Difficult or Delayed Activation (2577)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/01/2015 |
Event Type
malfunction
|
Event Description
|
It was reported that during pre-surgery, it was discovered that the small battery drive device would not activate.There were no delays to the planned surgical procedure.It was not reported if a spare device was available for use.There was no patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown; however, the reporter stated that the event occurred sometime during the week of (b)(6) 2015.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the device was not functional and did not engage when power was applied.Therefore the reported condition was confirmed.It was noted that the wires on the electronic control unit were damaged and the electric motor did not function.The assignable root cause was determined to be due to wear on electronic components from repeated sterilization and normal use over time.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|