SYNTHES OBERDORF BATTERY REAMER/DRILL II HANDPIECE; MOTOR, SURGICAL INSTRUMENT, AC-POWERED
|
Back to Search Results |
|
Catalog Number 530.705 |
Device Problem
Failure to Shut Off (2939)
|
Patient Problem
No Information (3190)
|
Event Type
malfunction
|
Event Description
|
It was reported from (b)(6) that the battery reamer/drill device could not be turned off.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
(b)(6).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If additional information should become available, a supplemental medwatch report will be sent accordingly.
|
|
Manufacturer Narrative
|
Additional narrative: the actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the motor power was too low and the trigger jams.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to false and defective material.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|