Model Number 5386
Device Problems Difficult to Interrogate (1331); Telemetry Discrepancy (1629)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  Injury  
Event Description
It was reported that during a follow up, interrogation of the pulse generator was difficult.When attempting to perform tests, loss of telemetry occurred.In between testing, there were no telemetry issues.Despite changing programmers and repositioning the telemetry wand, the issue remained.The device was explanted and replaced.There were no adverse consequences for the patient.
 
Manufacturer Narrative
(b)(4).Final analysis found that the pulse generator exhibited intermittent telemetry due to a discrepant integrated circuit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IDENTITY ADX XL DR
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC. (CRM-KISTA)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC. (CRM-KISTA)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
ulla strindlund
box 7051
isafjordsgatan 15
kista SE-16-407
SW   SE-16407
4684744043
MDR Report Key4902840
MDR Text Key6067213
Report Number3010215456-2015-27954
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeSW
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 05/21/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2004
Device Model Number5386
Device Lot Number0002346510
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/21/2015
Initial Date FDA Received07/09/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/29/2003
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-