The facility reported the thermometer reading indicated a temperature greater than 100 degrees repeatedly which resulted in diagnostic testing being ordered, iv antibiotics administered and the hospital stay extended.They later determined the elevated temperature reading was not correct.No sample was returned for eval.We have had no response to multiple requests asking for additional info pertaining to this report.We have not confirmed the readings were inaccurate of identified a root cause.The existence of other clinical symptoms and/or contributing factors is not known.However, due to the reported incident and in an abundance of caution, this medwatch is being filed.
|