MAUDE Adverse Event Report: BAXTER HEALTHCARE - SWINFORD CONNECTION SHIELD; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - SWINFORD CONNECTION SHIELD; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE Back to Search Results
Catalog Number SPC4211
Device Problems Break (1069); Leak/Splash (1354)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device was received for evaluation.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Visual inspection found the pouch to be pierced and verified the reported condition.The cause of the reported issue could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the cover of the iodine cap of a clamshell was broken.The cover came up from the overpouch and iodine leaked.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONNECTION SHIELD
Type of Device
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
BAXTER HEALTHCARE - SWINFORD
foxford road
swinford
EI 
Manufacturer (Section G)
BAXTER HEALTHCARE - SWINFORD
foxford road
swinford
EI  
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key4906345
MDR Text Key13511853
Report Number1416980-2015-28397
Device Sequence Number1
Product Code KDJ
Combination Product (y/n)N
Reporter Country CodeFI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 06/17/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date11/30/2016
Device Catalogue NumberSPC4211
Device Lot Number14L22H12
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/13/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/17/2015
Initial Date FDA Received07/10/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received07/23/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-