(b)(4) - the reported issue that the lead could not be fully inserted into the lower port was confirmed.As received, the ipg was responsive and communicated with lab utilities.Microscopic inspection identified material in the connector block.This obstruction prevented a lab lead from fully inserted into the port.This anomaly is consistent with silicone from inside the header assembly getting lodged in the connector block while trying to engage the setscrew.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.(b)(4).
|