Catalog Number PIP30
Device Problem Break (1069)
Patient Problem Failure of Implant (1924)
Event Date 06/24/2015
Event Type  Injury  
Event Description
It was reported that the patient had a pip implant break.The implant was in for less than 3 months.Sales rep stated that patient went in for follow up and stated they were feeling great and had good movement and upon x-ray it was discovered the implant was broken.
 
Manufacturer Narrative
Device will not be returned.If additional information becomes available it will be provided on a supplemental report.The reported device was manufactured and distributed by (b)(4) and implanted prior to howmedica osteonics corp.¿s purchase of certain assets of sbi on (b)(6) 2014.Stryker became legal manufacturer of this product on (b)(6) 2015 and has taken the responsibility for medical device reporting.Device will not be returned.
 
Manufacturer Narrative
The reported event that a 30 silicone pip implant (sterile packed) was alleged of breakage after surgery could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.Based on investigation, the root cause was attributed to a patient related issue.The most likely cause is excessive physical activity and/or abuse.Strenuous loading, excessive mobility, and articular instability all may lead to accelerated wear and eventual failure by fracture of the device.Patients should be made aware of the increased potential for device failure if excessive demands are made upon it.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Event Description
It was reported that the patient had a pip implant break.The implant was in for less than 3 months.Sales rep stated that patient went in for follow up and stated they were feeling great and had good movement and upon x-ray it was discovered the implant was broken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
30 SILICONE PIP IMPLANT (STERILE PACKED)
Type of Device
PROSTHESIS, FINGER, CONSTRAINED, POLYMER23
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
CH  2545
Manufacturer (Section G)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
CH   2545
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key4915768
MDR Text Key6008986
Report Number0008031020-2015-00304
Device Sequence Number1
Product Code KYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K931588
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,health professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/24/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberPIP30
Device Lot Number23987001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/24/2015
Initial Date FDA Received07/15/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received11/16/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/30/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-