AGA MEDICAL CORPORATION AMPLATZER SEPTAL OCCLUDER; CARDIAC OCCLUSION DEVICE Back to Search Results
Model Number 9-ASD-024
Device Problems Migration or Expulsion of Device (1395); Device Dislodged or Dislocated (2923)
Patient Problem Embolism (1829)
Event Date 06/25/2015
Event Type  Injury  
Event Description
A 24 mm amplatzer septal occluder (aso) was successfully placed in a very large atrial septal defect with little septal rim.Post-procedure before the patient was transferred from the cath lab, the 24 mm aso was found to have embolized into the right ventricle by fluoroscopy.The aso was percutaneously retrieved with no replacement.
 
Manufacturer Narrative
(b)(4).
 
Manufacturer Narrative
(b)(4) - congenital defect/deformity.Unintended movement.The results of this investigation confirmed the aso met all functional and dimensional specifications when analyzed at sjm.A review of the device history record confirmed the aso met all visual, dimensional, and functional specifications at the time it was manufactured, prior to shipment.There was no evidence to suggest there was an intrinsic defect in the aso, and the cause for the reported event remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Type of Device
CARDIAC OCCLUSION DEVICE
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key4922190
MDR Text Key19987774
Report Number2135147-2015-00078
Device Sequence Number1
Product Code MLV
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P000039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/26/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/17/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date12/31/2018
Device Model Number9-ASD-024
Device Catalogue Number9-ASD-024
Device Lot Number4517398
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/02/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/26/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/10/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Congenital Anomaly; Hospitalization; Required Intervention;
Patient Age4 YR
Patient Weight16
-
-