Two microsensors were returned and evaluated by the supplier.The supplier's evaluation included a review of quality records, which found that the devices met all manufacturing and quality testing/inspection specifications prior to distribution.The evaluation of the device with serial number (b)(4) found that there was no visible damage to the sensor and the sensor was functional.There was a cut in the catheter material 28 cm from the tip of the sensor case.Internal wires were exposed at the cut site.Due to the condition of the catheter as it was received, no further testing was possible.The supplier was unable to confirm the issue reported against this device.The condition of the catheter was determined to the be result of user handling.Based on the results of this investigation no further action is required.Trends will be monitored for this and similar complaints.At the present time this complaint is closed.Refer to mdrs: 1226348-2015-10428, 1226348-2015-10413.For information regarding additional devices reported in this event.
|