Catalog Number 06437281190
Device Problem High Test Results (2457)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/03/2015
Event Type  malfunction  
Manufacturer Narrative
As part of the investigation, the sample was tested on a centaur analyzer.The tsh results were below the normal reference range and ft4 and ft3 results were above the normal reference ranges.Based on this data, it was assumed either there was an issue with the sample such as hemolysis, lipemia, or contamination or the results received for the sample were correct.A general reagent issue was excluded.
 
Event Description
The customer received questionable thyroid results for one patient sample from a cobas e601 analyzer on an unknown date.The sample was repeated on an architect analyzer and the results for free thyroxine (ft4) and free triiodothyronine (ft3) were discrepant.Refer to the attachment to the medwatch for all patient data.(b)(4).Information concerning if any erroneous result was reported outside the laboratory was requested, but was not provided.No adverse event was reported.
 
Manufacturer Narrative
This event occurred in (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FT4, FREE THYROXINE
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 6830 5
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key4925774
MDR Text Key14378769
Report Number1823260-2015-03828
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K961489
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 07/20/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number06437281190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Event Location Laboratory
Initial Date Manufacturer Received 07/03/2015
Initial Date FDA Received07/20/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received07/20/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-