It was reported to covidien on (b)(6) 2015 that a customer had an issue with a dialysis catheter.The customer reports the patient was on dialysis approximately 5 minutes, blood was noted on the gauze dressing.There was visual confirmation of a pinhole, leaking blood from the venous lumen (pigtail).There was no patient injury.The femoral catheter was exchanged.The catheter was originally implanted on (b)(6) 2015.The patient current condition is stable.
|
The device history record was reviewed and indicated that the product was released accomplishing all quality standards.A sample was received for evaluation.There were no issues found during a visual inspection.The sample was submitted to an underwater leak test.Bubbles were detected coming from the venous extension lumen.The arterial extension lumen did not show bubbles.As per the instructions for use, it is necessary to perform a visual inspection before using the device.Do not use the catheter if it is crushed, cracked, cut or otherwise damaged.Clamping the catheter repeatedly in the same spot could weaken the tubing.Exercise caution when using sharp instruments near the catheter.The catheter tubing can tear when subjected to excessive force or rough edges.Inspect the catheter frequently for nicks, scrapes or cuts which could impair its performance.A possible root cause can be due to repeated clamping or the catheter may have come into contact with a sharp object.It must be noted that in-process controls (such as personnel training, incoming quality acceptance testing for raw material, 100% in process visual inspection, leak testing and visual acceptance sampling) are in place to prevent non-conforming product from leaving the manufacturing operations.As per procedure, manufacturing performs 100% leak testing and a 100% visual inspection at the final stage of production, which would identify this issue in the catheter assembly.No additional actions are required.This complaint will be used for tracking and trending purposes.
|