COVIDIEN PALINDROME 33/50 SPORT W/ SLOT; DIALYSIS CATHETER Back to Search Results
Model Number 8888133500
Device Problems Fluid/Blood Leak (1250); Hole In Material (1293)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/10/2015
Event Type  malfunction  
Manufacturer Narrative
Submit date: (b)(6) 2015.An investigation is currently under way; upon completion the results will be forwarded.
 
Event Description
It was reported to covidien on (b)(6) 2015 that a customer had an issue with a dialysis catheter.The customer reports the patient was on dialysis approximately 5 minutes, blood was noted on the gauze dressing.There was visual confirmation of a pinhole, leaking blood from the venous lumen (pigtail).There was no patient injury.The femoral catheter was exchanged.The catheter was originally implanted on (b)(6) 2015.The patient current condition is stable.
 
Manufacturer Narrative
The device history record was reviewed and indicated that the product was released accomplishing all quality standards.A sample was received for evaluation.There were no issues found during a visual inspection.The sample was submitted to an underwater leak test.Bubbles were detected coming from the venous extension lumen.The arterial extension lumen did not show bubbles.As per the instructions for use, it is necessary to perform a visual inspection before using the device.Do not use the catheter if it is crushed, cracked, cut or otherwise damaged.Clamping the catheter repeatedly in the same spot could weaken the tubing.Exercise caution when using sharp instruments near the catheter.The catheter tubing can tear when subjected to excessive force or rough edges.Inspect the catheter frequently for nicks, scrapes or cuts which could impair its performance.A possible root cause can be due to repeated clamping or the catheter may have come into contact with a sharp object.It must be noted that in-process controls (such as personnel training, incoming quality acceptance testing for raw material, 100% in process visual inspection, leak testing and visual acceptance sampling) are in place to prevent non-conforming product from leaving the manufacturing operations.As per procedure, manufacturing performs 100% leak testing and a 100% visual inspection at the final stage of production, which would identify this issue in the catheter assembly.No additional actions are required.This complaint will be used for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PALINDROME 33/50 SPORT W/ SLOT
Type of Device
DIALYSIS CATHETER
Manufacturer (Section D)
COVIDIEN
covidien manufacturing solulfons sa
edificio 820 calle #2 zona franca coyol
alajuela
CS 
Manufacturer (Section G)
COVIDIEN
covidien manufacturing solulfons sa
edificio 820 calle #2 zona franca coyol
alajuela
CS  
Manufacturer Contact
thom mcnamara
15 hampshire street
mansfield, MA 02048
5084524811
MDR Report Key4933193
MDR Text Key22657741
Report Number3009211636-2015-00314
Device Sequence Number1
Product Code MSD
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Followup
Report Date 07/14/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8888133500
Device Catalogue Number8888133500
Device Lot Number308D39X
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received07/22/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received10/24/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age84 YR
Patient Weight32
-
-