Catalog Number 5100060001 |
Device Problems
Use of Device Problem (1670); Failure to Auto Stop (2938); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/30/2015 |
Event Type
malfunction
|
Event Description
|
It was reported that during a cranial procedure, the perforator continued to run on longer than expected.It was further reported that there were no adverse consequence for the patient or delays to surgery as a result of the reported event.It was also reported that the surgery was completed successfully.
|
|
Manufacturer Narrative
|
Awaiting device evaluation.A follow up report will be filed once the quality investigation is complete.
|
|
Manufacturer Narrative
|
The device was returned for evaluation and functional testing of the disengagement mechanism confirmed that the mechanism functioned as intended.The definitive root cause could not be determined and potentially failure to follow instructions caused or contributed to this event.
|
|
Event Description
|
It was reported that during a cranial procedure, the perforator continued to run on longer than expected.It was further reported that there were no adverse consequence for the patient or delays to surgery as a result of the reported event.It was also reported that the surgery was completed successfully.
|
|
Search Alerts/Recalls
|