Catalog Number 05.001.201 |
Device Problem
Failure to Shut Off (2939)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Initial reporter¿s phone number: (b)(6).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If additional information should become available, a supplemental medwatch report will be sent accordingly.
|
|
Event Description
|
It was reported from (b)(6) that the battery handpiece device kept running after the device was switched off while using the device controls.It was not reported if the device was used in surgery, or if there was patient involvement.There were no delays in the surgical procedure.A spare device was not available for use.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the device had a hanging trigger.Therefore, the reported condition was confirmed.The assignable root cause was determined to be wear from normal use and servicing.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|