DEPUY SYNTHES POWER TOOLS SMALL BATTERY DRIVE (COLIBRI); INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESSORIES AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 532.001 |
Device Problems
Physical Resistance (2578); Mechanical Jam (2983)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the motor seized, was jammed and heavy moving.Therefore, the reported condition was confirmed.It was further determined that the motor failed and the control unit was defective.The assignable root cause was determined to be due to wear from normal use and servicing.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported from (b)(6) that before surgery, it was observed that the small battery drive device would hang up from time to time.During in-house engineering evaluation, it was observed that the motor seized, was jammed and heavy moving.It was further reported that the motor failed and the control unit was defective.It was not reported if there were any delays to the scheduled surgical procedure or if a spare device was available for use.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|