• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS FT4, FREE THYROXINE; RADIOIMMUNOASSAY, FREE THYROXINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS FT4, FREE THYROXINE; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Catalog Number 06437281160
Device Problem High Test Results (2457)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/09/2015
Event Type  malfunction  
Event Description
The customer received questionable free thyroxine (ft4) results for one patient sample from cobas e601 serial number (b)(4).The initial result was 7.49 ng/dl and was reported outside the lab.The doctor questioned the result and the sample was repeated with a result of 7.63 ng/dl.The customer sent the sample to a reference lab where it was tested by chemiluminescense and the result was 1.25 ng/dl.The result from the reference laboratory was believed to be correct.The patient was not adversely affected.The field service representative determined the issue was with the patient sample.He ran performance testing.All controls were acceptable and the system was performing to specifications.
 
Manufacturer Narrative
It was unknown if the initial reporter sent report to the fda.
 
Manufacturer Narrative
Additional information was received that the analyzer used at the reference laboratory was an abbott architect.A specific root cause could not be identified as no sample was available for further investigation.Based on analysis of the provided calibration and qc data, a general reagent issue was excluded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FT4, FREE THYROXINE
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key4953377
MDR Text Key20040469
Report Number1823260-2015-03892
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K961489
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,health professional,use
Reporter Occupation Health Professional
Type of Report Initial
Report Date 08/21/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/30/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2016
Device Catalogue Number06437281160
Device Lot Number18347401
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Event Location Laboratory
Date Manufacturer Received07/15/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
LEVOTHYROXINE
-
-