NEUROPACE, INC. NEUROPACE RNS SYSTEM Back to Search Results
Model Number RNS NEUROSTIMULATOR
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/12/2013
Event Type  Injury  
Event Description
On (b)(6) 2013 - the patient had her third neurostimulator replacement procedure.The rns system configuration for this subject includes a rns neurostimulator, one cortical strip lead (connected, cl-315-10) a depth lead (connected, dl-330-10) and a depth lead (dl-344-10, not connected).Prior to the replacement intraoperative cultures were taken and were positive for gram positive bacillus on gram stain.At time of surgery there were no overt signs of infection.Patient denies fevers, chills or sweats prior to surgery.Patient had no complaints of pain and noted no drainage from the incision site.Blood work obtained revealed a normal white blood count.The neurostimulator replacement procedure occurred (b)(6) 2013.
 
Manufacturer Narrative
(b)(4).On 5/12/2015 during the adverse event review for the pas study it was identified that this event had not been reviewed according to the complaint handling process.Once identified this record was generated to meet the complaint review process.Neuropace, inc.Voided this complaint.There was no reported product malfunction and no indication the product contributed to the infection.Infections are anticipated when performing craniotomy procedures.On 7/17/2015 neuropace, inc.Reevaluated this record.Based on internal review of the fda regulations, guidance document (draft guidance for industry and food and drug administration staff), and internal procedures, neuropace, inc.Determined that complaints associated with incision site infections are reportable.These past events show no indication that the product caused the infection and there is no indication of product failure.Sterility record review performed.No evidence of factors that could have caused or contributed to the reported event.All sterilization results met requirements as specified in the specifications.The patient is currently implanted with the rns system and is programmed for detection and therapy.Product remains implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROPACE RNS SYSTEM
Type of Device
NEUROPACE RNS SYSTEM
Manufacturer (Section D)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer (Section G)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer Contact
ramona gonis
455 n. bernardo ave.
mountain view, CA 94043
6502382788
MDR Report Key4954168
MDR Text Key18296462
Report Number3004426659-2015-00017
Device Sequence Number1
Product Code PFN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 07/17/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Model NumberRNS NEUROSTIMULATOR
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/12/2013
Initial Date FDA Received07/30/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age58 YR
-
-