Catalog Number 512.100
Device Problems Partial Blockage (1065); Physical Resistance (2578); Mechanical Jam (2983)
Patient Problem No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
(b)(6).The actual device was returned for evaluation with an unspecified malfunction.Reliability engineering evaluated the device and observed that the trigger/slider was blocked, jammed and heavy moving.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported from (b)(6) that the mini air drill device had an unspecified malfunction.During in-house engineering evaluation, it was observed that the trigger/slider was blocked, jammed and heavy moving.It was not reported if the device was used in surgery or if there was patient involvement reported.It was not reported if there any delays in the surgical procedure or if a spare device was available for use.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINI AIR DRILL
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESSORIES AND ATTACHMENT
Manufacturer (Section D)
SYNTHES BETTLACH
muracherstrasse 3
bettlach CH254 4
SZ  CH2544
Manufacturer (Section G)
SYNTHES BETTLACH
muracherstrasse 3
bettlach CH254 4
SZ   CH2544
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key4957807
MDR Text Key23233750
Report Number9612488-2015-10383
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Report Date 07/23/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/31/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number512.100
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/09/2014
Date Manufacturer Received07/14/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/06/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-