Model Number DTBA2D1
Device Problem Pacing Problem (1439)
Patient Problems Syncope (1610); Bruise/Contusion (1754)
Event Date 05/26/2015
Event Type  Injury  
Event Description
It was reported that the patient had syncope and the injury of a bruise due to at least two ventricular tachycardias (vt) terminating with single shocks coinciding with left ventricular (lv) lead capture management on their implantable cardioverter defibrillator (icd).The device was reprogrammed.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
This model number is not approved for distribution in the united states; however, it is same/similar to a device marketed in the u.S.This event occurred outside the us and patient information is not generally available due to confidentiality concerns.(b)(4).
 
Manufacturer Narrative
Product event summary: the device was not returned for analysis.However, performance data collected from the device was received and analyzed.Analysis of the device memory indicated that during the capture management feature running, ventricular tachycardia was induced and terminated with a single shock therapy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIVA XT
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
IPG MFG SWITZERLAND
route du molliau 31
tolochenaz 1131
CH  1131
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key4967624
MDR Text Key6637734
Report Number9614453-2015-01998
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,company representati
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/01/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/04/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/14/2015
Device Model NumberDTBA2D1
Device Catalogue NumberDTBA2D1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/01/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/27/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Required Intervention;
Patient Age00070 YR
-
-