• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. VIVA QUAD XT; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. VIVA QUAD XT; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO Back to Search Results
Model Number DTBA1QQ
Device Problems False Alarm (1013); Incorrect Measurement (1383)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/08/2014
Event Type  malfunction  
Manufacturer Narrative
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.Product event summary: the actual product was not returned for analysis.However, performance data was collected from the device and was analyzed.Analysis of the device memory was performed and no anomalies were found.No anomalies identified.(b)(4).
 
Event Description
It was reported that the implantable cardioverter defibrillator (icd) device triggered warnings for all lead impedances, but displayed date of the warning was for a period before the implant of the device.Device data was reviewed and revealed that there was low battery voltage due to battery conditioning charge.The device remains in use.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIVA QUAD XT
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key4968581
MDR Text Key23533770
Report Number3004209178-2015-15067
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Nurse
Report Date 05/29/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/04/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/14/2016
Device Model NumberDTBA1QQ
Device Catalogue NumberDTBA1QQ
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/29/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/15/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age00068 YR
Patient Weight53
-
-