• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES USA; NAIL, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES USA; NAIL, FIXATION, BONE Back to Search Results
Device Problem Failure to Align (2522)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/23/2015
Event Type  malfunction  
Manufacturer Narrative
Patient information is not available for reporting.This report is for one (1) unknown proximal humeral nail.Based upon the part family for proximal humeral nails, a common name of hsb ¿ rod, fixation, intramedullary ¿ is also applicable.Device was not explanted.Investigation could not be completed and no conclusion could be drawn as no device was returned.Without a lot number, the device history record review could not be requested.Device used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes (b)(4) reports an event in (b)(6) as follows: it was reported that a proximal humeral fracture surgery was carried out with the multiloc humeral nailing system (short) on (b)(6) 2015.After the proximal region was fixed, the surgeon drilled the distal side for distal locking.However, the drill missed the target nail.Then, the surgeon held a sleeve and drilled again, putting the weight backward.Eventually, the drill successfully passed through the nail.The procedure was extended for fifteen (15) minutes.This report is for one (1) unknown proximal humeral nail.This report is 2 of 3 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
NAIL, FIXATION, BONE
Manufacturer (Section D)
SYNTHES USA
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
SYNTHES USA
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key4997344
MDR Text Key22859782
Report Number2520274-2015-15163
Device Sequence Number1
Product Code JDS
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Report Date 07/23/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/12/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-