DEPUY SYNTHES POWER TOOLS OSCILL-SAW-ATTACHM W/QUICK-COUPL F/COMPA; BLADE, SAW, GENERAL AND PLASTIC SURGERY, SURGICAL Back to Search Results
Catalog Number 511.801
Device Problem Connection Problem (2900)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(6).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported from (b)(6) that the attachment device could not hold the blade device properly and did not turn smoothly.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
The actual device was returned for evaluation.Reliability engineering evaluated the device and the reported condition that the device cannot hold properly the blade and it does not turn smoothly was confirmed.An assessment was performed and found that the coupling tool side was not functioning and was defective.It was further noted that the clamping screw was broken and defective.The assignable root cause was determined to be due to wear from normal use and servicing over time.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSCILL-SAW-ATTACHM W/QUICK-COUPL F/COMPA
Type of Device
BLADE, SAW, GENERAL AND PLASTIC SURGERY, SURGICAL
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ  CH4437
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ   CH4437
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key4997846
MDR Text Key24065625
Report Number3009450871-2015-12112
Device Sequence Number1
Product Code GFA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Report Date 05/20/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/12/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number511.801
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/19/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received09/01/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/12/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-