SORIN GROUP DEUTSCHLAND S5 DOUBLE HEAD PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-85-00
Device Problems Device Inoperable (1663); Other (for use when an appropriate device code cannot be identified) (2203)
Patient Problem No Patient Involvement (2645)
Event Date 07/07/2015
Event Type  malfunction  
Manufacturer Narrative
The serial number was not provided.The manufacturing date cannot be determined.The information will be forwarded if available.Sorin group (b)(4) manufactures the s5 double head pump.The incident occurred in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).The investigation is ongoing.A follow up report will be sent when the investigation is complete.
 
Event Description
Sorin group (b)(4) received a report that the lock function of the double head pump was not functioning during maintenance/repair.There was no patient involvement.
 
Manufacturer Narrative
The initial report indicated that the serial number was unknown.This is incorrect; the serial number is (b)(4).Device manufacture date (mm/dd/yyyy): 10/31/2014.(b)(4) manufactures the s5 double head pump.The incident occurred in (b)(6).This medwatch report is filed on behalf of (b)(4).A (b)(4) field service representative was dispatched to the facility to investigate.The service representative was able to confirm the reported issue and traced the failure back to locking pin that had fallen off.The pin was reattached to resolve the issue.The device was tested without further malfunction and was returned to service.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.Evaluated on site by (b)(4) technician.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 DOUBLE HEAD PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindbergh strasse 25
munich D-809 39
GM  D-80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key5001249
MDR Text Key24798977
Report Number9611109-2015-00271
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Unknown
Remedial Action Other
Type of Report Followup
Report Date 03/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/08/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-85-00
Was Device Available for Evaluation? Yes
Date Manufacturer Received02/07/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/31/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-