Model Number 10-85-00 |
Device Problems
Device Inoperable (1663); Other (for use when an appropriate device code cannot be identified) (2203)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 07/07/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
The serial number was not provided.The manufacturing date cannot be determined.The information will be forwarded if available.Sorin group (b)(4) manufactures the s5 double head pump.The incident occurred in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).The investigation is ongoing.A follow up report will be sent when the investigation is complete.
|
|
Event Description
|
Sorin group (b)(4) received a report that the lock function of the double head pump was not functioning during maintenance/repair.There was no patient involvement.
|
|
Manufacturer Narrative
|
The initial report indicated that the serial number was unknown.This is incorrect; the serial number is (b)(4).Device manufacture date (mm/dd/yyyy): 10/31/2014.(b)(4) manufactures the s5 double head pump.The incident occurred in (b)(6).This medwatch report is filed on behalf of (b)(4).A (b)(4) field service representative was dispatched to the facility to investigate.The service representative was able to confirm the reported issue and traced the failure back to locking pin that had fallen off.The pin was reattached to resolve the issue.The device was tested without further malfunction and was returned to service.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.Evaluated on site by (b)(4) technician.
|
|