• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP ARCTIC FRONT ADVANCE CARDIAC CRYOABLATION CATHETER; PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP ARCTIC FRONT ADVANCE CARDIAC CRYOABLATION CATHETER; PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Model Number 2AF284
Device Problem Insufficient Information (3190)
Patient Problem Pericardial Effusion (3271)
Event Date 07/16/2015
Event Type  Injury  
Manufacturer Narrative
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.Investigation is in progress.Results of investigation will be submitted in a supplemental report.(b)(4).
 
Event Description
It was reported that during a cryoablation procedure a moderately sized pericardial effusion was observed requiring a transthoracic echocardiogram and critical care monitoring postoperatively with extended an hospital stay.No pericardiocentesis was noted.The case was able to be completed with cryo.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: data files were returned and analyzed.Data files showed a system notice was received indicating that the refrigerant delivery path was obstructed, which is unrelated to the reported clinical issue encountered during the procedure.No product malfunction was reported and no product was returned for analysis.The reported adverse event could not be confirmed.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTIC FRONT ADVANCE CARDIAC CRYOABLATION CATHETER
Type of Device
PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
16771 chemin ste-marie
kirkland,qc H9H 5 H3
CA  H9H 5H3
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key5001395
MDR Text Key22898692
Report Number3002648230-2015-00232
Device Sequence Number1
Product Code OAE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100010/S015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Followup
Report Date 07/16/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator No Information
Device Expiration Date03/01/2016
Device Model Number2AF284
Device Catalogue Number2AF284
Device Lot Number95599
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/13/2015
Supplement Dates Manufacturer Received09/14/2015
Supplement Dates FDA Received09/18/2017
Was Device Evaluated by Manufacturer? No
Date Device Manufactured03/01/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening;
-
-