A us distributor contacted zoll to report that a (b)(6) patient was treated and passed away while in the icu on (b)(6) 2015.The lifevest detected ventricular fibrillation (vf) and delivered a 150j pulse at 23:01:23.The arrhythmia was converted to bradycardia.A second treatment was delivered at 23:14:21 during sinus tachycardia with bbb.Elevated heart rate with bbb contributed to the false detection.The patient's post-shock rhythm remained sinus tachycardia.The lifevest detected cpr artifact after the second treatment, and the response buttons were pressed.The patient was externally defibrillated in the icu at 23:28:25 during a run of vf.Cpr artifact interrupted the lifevest vf detection.The patient's post-shock rhythm could not be determined due to cpr artifact.The electrode belt was disconnected at 23:35:36 and the patient subsequently passed away.
|