DEPUY SYNTHES POWER TOOLS COLIBRI; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 532.001 |
Device Problems
Mechanical Problem (1384); Power Conditioning Problem (1474); Physical Resistance (2578); Device Operates Differently Than Expected (2913); Mechanical Jam (2983)
|
Patient Problem
No Information (3190)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Additional narrative: (b)(6).The actual device was returned for evaluation.Reliability engineering evaluated the device and the reported condition was confirmed.An assessment was performed and found that the motor had seized, was jammed, heavy moving and was rough running.It was further noted that the motor failed.The assignable root cause was determined to be due to wear from normal use and servicing over time.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Event Description
|
It was reported from (b)(6) that the small battery drive device had low power.During in-house engineering evaluation, it was found that the motor had seized, failed, was jammed, heavy moving and was rough running.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|