• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL,INC. ENDOWRIST ONE VESSEL SEALER; ENDOSCOPIC ELECTROSURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL,INC. ENDOWRIST ONE VESSEL SEALER; ENDOSCOPIC ELECTROSURGICAL INSTRUMENT Back to Search Results
Model Number 410322-05
Device Problem Device Stops Intermittently (1599)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/22/2015
Event Type  malfunction  
Manufacturer Narrative
The endowrist one vessel sealer instrument has not been returned for evaluation, therefore, the root cause of the customer reported failure mode cannot be determined.A follow-up mdr will be submitted if the instrument is returned (post engineering evaluation) or if additional information is received.Based on the information provided, this complaint is being reported due to the following conclusion: during a da vinci distal pancreatectomy procedure, allegedly, the endowrist one vessel sealer instrument intermittently cauterized patient's tissue.While there was no harm to the patient; recurrence of the alleged failure mode could likely cause or contribute to an adverse event if the failure mode were to recur.Review of the site's system logs for the reported procedure date found that no system errors were generated that would have caused or contributed to the reported issue.
 
Event Description
It was reported that during a da vinci surgical procedure, the endowrist one vessel sealer instrument worked intermittently and cauterized sometimes.The planned surgical procedure was completed and there was no report of any patient harm or adverse outcome.On 07/30/2015, intuitive surgical, inc.(isi) received additional information from the isi clinical sales representative (csr) who initially reported this complaint.According to the csr, the planned surgical procedure was a da vinci distal pancreatectomy procedure.The csr does not recall the tissue that was being sealed; however, during the sealing cycle audible tone was heard indicating that the sealing cycle was completed.The csr indicated that no error messages occurred during the sealing cycle.Tissue bunching in the jaws of the instrument was not observed.The csr indicated that the surgeon believes that the vessel sealer instrument failed to properly cauterize.No other information regarding the reported event was provided.
 
Manufacturer Narrative
Intuitive surgical, inc.Had performed a review of the site's system log and it was found that the reported endowrist one vessel sealer instrument was not used on 04/22/2015, which was previously reported in the initial mdr.At this time, the actual date of when the reported event occurred with the reported instrument is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST ONE VESSEL SEALER
Type of Device
ENDOSCOPIC ELECTROSURGICAL INSTRUMENT
Manufacturer (Section D)
INTUITIVE SURGICAL,INC.
sunnyvale CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC.
950 kifer rd.
sunnyvale CA
Manufacturer Contact
tabitha reed
950 kifer rd.
sunnyvale, CA 
MDR Report Key5033934
MDR Text Key25401695
Report Number2955842-2015-01139
Device Sequence Number1
Product Code NAY
Combination Product (y/n)N
PMA/PMN Number
K110639
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Report Date 07/30/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number410322-05
Device Lot NumberM11141208
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Event Location Hospital
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/26/2015
Was Device Evaluated by Manufacturer? No
Date Device Manufactured12/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-