• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECKMAN COULTER COULTER AC T DIFF 2 ANALYZER; COUNTER, DIFFERENTIAL CELL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECKMAN COULTER COULTER AC T DIFF 2 ANALYZER; COUNTER, DIFFERENTIAL CELL Back to Search Results
Catalog Number 6605500
Device Problem Contamination (1120)
Patient Problem Not Applicable (3189)
Event Date 08/06/2015
Event Type  malfunction  
Manufacturer Narrative
The field service engineer (fse) evaluated the instrument on 08/06/2015.The fse found debris in the white blood cell, wbc, aperture and a dirty probe wipe.He replaced the wbc bath and the instrument ran without any hemoglobin, hgb, failures.He also replaced the probe wipe to address the dirty probe wipe.The repairs were verified per established service procedures.(b)(6).
 
Event Description
The customer observed hgb failing background from a coulter ac t diff 2 analyzer while performing startup.Erroneous patient results were not generated and there was no change or affect to patient treatment in connection with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COULTER AC T DIFF 2 ANALYZER
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s kraemer blvd
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th avenue
kendall FL 33196
Manufacturer Contact
miranda holland
11800 sw 147th avenue
32-s08
kendall, FL 33196
3053802031
MDR Report Key5033959
MDR Text Key24202653
Report Number1061932-2015-01296
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K990352
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Report Date 08/06/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/27/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number6605500
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/06/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-