The customer experienced an issue with qc and patient results for hemoglobin a1c on integra 800 analyzer serial number (b)(4) over several days.The customer stated the patient results "are swinging wildly as much as 6, as in going from a 7 to a 13%".Around 500 patient samples were involved.Of the data provided for 75 patient samples, only the results for 15 were discrepant.The following data is initial result, followed by repeat result on 08/11/2015.Patient 1: 8.6%, 5.3%.Patient 2: 6%, 4.0%.Patient 3: 6.1%, 7.7%.Patient 4: 6.2%, 7.9%.Patient 5: 6.8%, 8.5%.Patient 6: 6.1%, 7.8%.Patient 7: 6.2%, 8.1%.Patient 8: 7%, 9.0%.Patient 9: 7.3%, 9.4%.Patient 10: 7.5%, 9.7%.Patient 11: 7.4%, 9.6%.Patient 12: 7.9%, 10.2%.Patient 13: 8.1%, 10.6%.Patient 14: 8.4%, 10.9%.Patient 15: 9.4%, 13.2%.All of the initial results were reported outside the laboratory.The repeat results were believed to be correct.Some patients were notified that they would be followed as diabetic based on the original values.Corrective reports and notifications were being sent.The patients were not adversely affected.The customer believed the issue was reagent pack related.The field service representative noted that when the shipment with the suspect reagent was received on 07/14/2015, it was very hot and humid locally.The field service representative checked the analyzer and could not determine a cause.He replaced the air dryers, checked the rotor light barrier adjustment, and ran a work station accuracy check which passed.He checked the calibration history and the customer ran calibration and qc which passed.
|