• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET ORTHOPEDICS PRELUDE PF THD GUIDE PIN 5PK; TEMPLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET ORTHOPEDICS PRELUDE PF THD GUIDE PIN 5PK; TEMPLATE Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 05/13/2015
Event Type  Injury  
Manufacturer Narrative
Current information is insufficient to permit a conclusion as to the cause of the event.Review of device history records show that lot released with no recorded anomaly or deviation.There are warnings in the package insert that state that this type of event can occur: under precautions states, ¿intraoperative fracture or breaking of instruments has been reported.¿.
 
Event Description
It was reported patient underwent a patellofemoral replacement procedure that utilized guide pins on (b)(6) 2015.Post-operative radiographs taken two weeks after surgery revealed that a fractured piece of the guide pin was retained by the patient.The surgeon was unaware that the guide pin fractured during the initial procedure.There is no revision scheduled at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRELUDE PF THD GUIDE PIN 5PK
Type of Device
TEMPLATE
Manufacturer (Section D)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
megan haas
56 e. bell drive
warsaw, IN 46582
5743726700
MDR Report Key5036687
MDR Text Key24353999
Report Number0001825034-2015-03796
Device Sequence Number1
Product Code HWT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Report Date 08/13/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Model NumberN/A
Device Catalogue Number110004170
Device Lot Number791810
Other Device ID NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/27/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age51 YR
-
-