PHYSIO-CONTROL, INC HANDLES AND CONNECTOR WITH DISCHARGE CONTROL FOR INTERNAL; DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES) Back to Search Results
Catalog Number 3006569
Device Problem Failure to Deliver Energy (1211)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/03/2015
Event Type  malfunction  
Manufacturer Narrative
Physio-control evaluated the internal defibrillation paddles and was unable to verify or duplicate the reported failure to deliver defibrillation energy.The cause of the reported issue could not be determined.
 
Event Description
The customer reported that during a patient event, the internal defibrillation paddles did not deliver defibrillation energy.The customer indicated that they attempted defibrillation delivery 3 or 4 times and then was able to use a back up set of internal defibrillation paddles successfully.The patient did not suffer any adverse outcome during the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDLES AND CONNECTOR WITH DISCHARGE CONTROL FOR INTERNAL
Type of Device
DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES)
Manufacturer (Section D)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 9706
Manufacturer (Section G)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 9706
Manufacturer Contact
jason march
4258674000
MDR Report Key5040382
MDR Text Key24581303
Report Number3015876-2015-01050
Device Sequence Number1
Product Code LDD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K895379
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Biomedical Engineer
Report Date 08/28/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number3006569
Device Lot Number0105
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2015
Is the Reporter a Health Professional? Yes
Event Location Hospital
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/28/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-