Hill-rom technical support provided the part number for the utv board.Per the hill-rom service manual the advanta bed requires an effective maintenance program.Preventative maintenance will minimize downtime due to excessive wear.Communications: inspect and test the communication junction box.Test the sidecom communication system features for proper operation.Inspect the communication cable including the male and female pins in the plug.Test the nurse call super capacitor for proper operation.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The account replaced the utv board to resolve the issue.Based on this information, no further action is required.
|