It was reported that during use of the vigilance monitor, the cco value was abnormal - approximately 2l/min higher.However, ecg and map were as expected.No error messages were reported and the customer was not performing sv02 measurement when the atypical value was observed.There was no report of patient compromise or inappropriate treatment, and no additional system-related devices were initially identified as suspect.However, when the vigilance monitor was returned for evaluation, a 70cc2 cable was discovered attached to the monitor.While the 70cc2 cable was not implicated in the customer's report, an additional complaint was initiated to consider the cable as suspect.
|