Model Number 2AF284 |
Device Problem
Insufficient Information (3190)
|
Patient Problems
Abdominal Pain (1685); Nausea (1970)
|
Event Date 08/05/2015 |
Event Type
Injury
|
Manufacturer Narrative
|
This event occurred outside the us where the same model is distributed.All information provided is included in this report.Patient information is not generally available due to confidentiality concerns.(b)(4).
|
|
Event Description
|
It was reported that the day after a cryo ablation procedure, the patient complained of nausea and gastric discomfort.An x-ray revealed that the patient had a dilated stomach.The patient was transferred to another hospital for specialty care.No further patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: bin files did not show any system notices for the date of procedure.Bin files also showed at least 8 injections were performed with catheter 2af284 / 57264-78.Historical data review showed no occurrences of the reported adverse event for catheter 2af284 with lot #57264.This was a post-procedure clinical event.No product malfunction reported.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|